EEPROM version newer than ROM version – (old software). Contact Fisher & Paykel. Model mismatch with software, contact your Fisher & Paykel Healthcare. NOTE: included in software version 7.14 onward, except 7.21. Temperature circuit calibration out of range: Range Amp 0: 25.5 °C. Neopuff™ / Perivent™ Operating Manual (Fisher & Paykel Healthcare Part No. 185041726) for set-up instructions. Unscrew PEEP cap for disinfection 2.4 Sterilization CAUTION:. Do not autoclave or gas-sterilize any part of the F&P Neopuff™ / Perivent™ Infant Resuscitator. Do not autoclave any of the Neopuff / Perivent accessories. Fisher And Paykel Humidifier Service Manual Download INTERNATIONAL P O Box 14 348, Panmure, Auckland 1134, New Zealand Tel:+64 9 574 0100 Fax:+64 9 574 0158 Email: info@fphcare.com Web Site: www.fphcare.com USA Tel: 1800 446 3908 or +1 949 470 3900 Fax:+1 949 470 3933.
Download 62 pages, 3.91 MbFisher Paykel Repair Manual
Fisher&paykel Airvo 2 Service Manual
6.3“See Manual” Error Codes
The following is an explanation of the error codes that are displayed in conjunction with the See Manual indicator flashing. A code is not displayed if the microprocessor has stopped functioning (see technical problems - section 6.2).
Error | Description of Fault | |
E00 | No fault | |
E02 | Microprocessor stack overflow | |
E03 | RAM fault | |
E04 | ROM fault | |
E05 | EEPROM version older than ROM version. Update EEPROM. Refer to Section 7.2.6 | |
E06 | EEPROM version newer than ROM version – (old software). Contact Fisher & Paykel | |
Healthcare | ||
E07 | Model mismatch with software, contact your Fisher & Paykel Healthcare | |
representative. | ||
NOTE: included in software version 7.14 onward, except 7.21. | ||
E10 | Temperature circuit calibration out of range: Range Amp 0: 25.5 °C | |
E11 | Temperature circuit calibration out of range: Range Amp 0: 65.0 °C | |
E12 | Temperature circuit calibration out of range: Range Amp 1: 25.5 °C | |
E13 | Temperature circuit calibration out of range: Range Amp 1: 34.5 °C | |
E14 | Temperature circuit calibration out of range: Range Amp 2: 34.5 °C | |
E15 | Temperature circuit calibration out of range: Range Amp 2: 44.6 °C | |
E16 | Temperature circuit calibration out of range: Range Amp 3: 65.0 °C | |
E20 | Heater wire circuit has malfunctioned, heater can not be energized: | |
- Heater plate thermal cutout tripped (section 7.2.5) | ||
- Heater wire fuse (F2) is open circuit (section 7.2.2) | ||
- Heater wire triac, or heater wire relay is open circuit | ||
E21 | Heater wire circuit has malfunctioned, heater either can not be | |
monitoring circuit has failed : | ||
- Heater wire triac has shorted | ||
- Heater wire sense circuit has failed | ||
E23 | Heater wire voltage measurement circuit faulty | |
E25 | Transistor Q17 is not turning on | |
E26 | Transistor Q16 is not turning on | |
E27 | The heater wire relay is short circuited | |
E28 | Mains voltage measurement is uncalibrated (section 8.1.3) | |
E29 | The heater plate thermistor is short circuited | |
E2A | The heater plate thermistor is open circuit | |
E2C | Heater plate element circuit has malfunctioned is not turning on: | |
- Element is open circuit | ||
- Heater plate triac or driver faulty | ||
- Heater plate relay is faulty | ||
E2D | The heater plate element is not turning off. Heater plate triac circuit faulty. | |
E30 | Power (on/off) button stuck on | |
E31 | Mute button stuck on | |
E32 | Mode button stuck on | |
E40 | Unit not functional tested at time of manufacture | |
MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) | 27 |
6.3“See Manual” Error Codes
The following is an explanation of the error codes that are displayed in conjunction with the See Manual indicator flashing. A code is not displayed if the microprocessor has stopped functioning (see technical problems - section 6.2).
Fisher Paykel Owners Manual
Error | Description of Fault | |
E00 | No fault | |
E02 | Microprocessor stack overflow | |
E03 | RAM fault | |
E04 | ROM fault | |
E05 | EEPROM version older than ROM version. Update EEPROM. Refer to Section 7.2.6 | |
E06 | EEPROM version newer than ROM version – (old software). Contact Fisher & Paykel | |
Healthcare | ||
E07 | Model mismatch with software, contact your Fisher & Paykel Healthcare | |
representative. | ||
NOTE: included in software version 7.14 onward, except 7.21. | ||
E10 | Temperature circuit calibration out of range: Range Amp 0: 25.5 °C | |
E11 | Temperature circuit calibration out of range: Range Amp 0: 65.0 °C | |
E12 | Temperature circuit calibration out of range: Range Amp 1: 25.5 °C | |
E13 | Temperature circuit calibration out of range: Range Amp 1: 34.5 °C | |
E14 | Temperature circuit calibration out of range: Range Amp 2: 34.5 °C | |
E15 | Temperature circuit calibration out of range: Range Amp 2: 44.6 °C | |
E16 | Temperature circuit calibration out of range: Range Amp 3: 65.0 °C | |
E20 | Heater wire circuit has malfunctioned, heater can not be energized: | |
- Heater plate thermal cutout tripped (section 7.2.5) | ||
- Heater wire fuse (F2) is open circuit (section 7.2.2) | ||
- Heater wire triac, or heater wire relay is open circuit | ||
E21 | Heater wire circuit has malfunctioned, heater either can not be | |
monitoring circuit has failed : | ||
- Heater wire triac has shorted | ||
- Heater wire sense circuit has failed | ||
E23 | Heater wire voltage measurement circuit faulty | |
E25 | Transistor Q17 is not turning on | |
E26 | Transistor Q16 is not turning on | |
E27 | The heater wire relay is short circuited | |
E28 | Mains voltage measurement is uncalibrated (section 8.1.3) | |
E29 | The heater plate thermistor is short circuited | |
E2A | The heater plate thermistor is open circuit | |
E2C | Heater plate element circuit has malfunctioned is not turning on: | |
- Element is open circuit | ||
- Heater plate triac or driver faulty | ||
- Heater plate relay is faulty | ||
E2D | The heater plate element is not turning off. Heater plate triac circuit faulty. | |
E30 | Power (on/off) button stuck on | |
E31 | Mute button stuck on | |
E32 | Mode button stuck on | |
E40 | Unit not functional tested at time of manufacture | |
MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) | 27 |